The CT
60 Length Gauge features a measuring range of 60 mm.
It is used for incorporation into special measuring apparatus,
calibrating and inspecting devices or special purpose
machines, for example in conjunction with the monitoring and
control of assembly and adjustment procedures.
A HEIDENHAIN-CERTO length gauging
system comprises:
CT 60 Length
Gauge,
ND 281 Display
Unit
Control box
Gauge stand
Ceramic suction
plate as support surface
Accuracy
The total error of the CT 60 over the entire measuring range
of 60 mm lies within ± 0.1 µm
at ambient temperatures between 19 and 21 °C and with
temperature fluctuations of ± 0.1 K during measurement.
When the CERTO is used with length error compensation, e.g.,
in conjunction with the ND
281Display Unit,
HEIDENHAIN guarantees a total error of
± 0.05 µm over the 60 mm
measuring range.
The CT
25 Length Gauge features a measuring range of 25 mm.
The high accuracy and measuring steps of 10 nm and less permit
use of the CT 25 in gauge block
calibration. Thanks to its generous measuring range —
much larger than with inductive length gauges — the number
of reference standard blocks required can be reduced
significantly.
Accuracy
The total error of the CT 25 over the entire measuring
range of 25 mm lies within ± 0.1 µm
at ambient temperatures between 19 and 21 °C and with
temperature fluctuations of ± 0.1 K during measurement.
When the CERTO is used with length error compensation, e.g.,
in conjunction with the ND
281Display Unit,
HEIDENHAIN guarantees a total error of
± 0.03 µm over the 25 mm
measuring range.