Call Us Toll Free 1 888 354 2525
surveillance
masters for calipers |
Vernier caliper masters are used for checking
outside jaws, inside jaws, and parallelism of jaws. Lapped gaging faces
are accurate with ± .0002" OR ± .005 MM from the nominal size.
Black-oxided to prevent rust.
Vernier Calipers, like all other types of measuring
equipment require frequent calibration on a regular, established inspection
cycle. This is true regardless of their relative sophistication (mechanical, or
dial types). This can be done using gage blocks or pins. The GSG Vernier Caliper
Master provides multiple checks of the caliper, which is necessary due to the
typical rack & gear design or glass scale design. The VC-1 inch provides
1 inch, through 6-inch checks
on the outside caliper along with ½ inch to 4-½ inch
on the inside caliper. These multiple checks verify
the caliper's accuracy throughout its range of us
e.
Other Caliper Masters |